
Zustand: Wiederaufbereitet durch MEDSER, hervorragender technischer und optischer Zustand
Standort: Installiert im CT-Cube
Röhre: Straton MX-P46
Verfügbarkeit: Bitte anfragen
| Aperture | 78 cm |
| Scan Field | 50 cm |
| Tilt | +/- 30° |
| Rotational times | 0.33, 0.5, 1.0 s |
| Number of detector rows | 32 |
| Elements | 23.552 |
| Channels per slice | 1472 |
| Number of projections | Up to 4.608 (1/360°) |
| Max. slices / rotation | 64 |
| Field of view | 50, 65 HD, 80 Extended cm |
| Reconstruction time | up to 40 images/s |
| Reconstruction slice widths | As low as 0.33 mm |
| Slice increment | 0.1–10 mm |
| Pitch factor (Volume Pitch) | Up to 1.5 |
| Scan length | max. 160 cm |
| Spiral scan time | Max. 100 s |
| Tube | STRATON |
| Tube current range | 20–800 mA |
| Tube voltages | 70, 80, 100, 120, 140 kV |
| Anode heat storage capacity | 50 MHU (cooling rate of 7.3 MHU/min) |
| Focal Spot size according to | 0.7 x 0.7 mm/7° |
| IEC 60336 | 0.9 x 1.1 mm/7° |
| Max. table load | 220 kg |
| Table speed | 2–200 mm/s |
| Vertical table travel range | 48–92 cm (at table top) |
| Vertical travel speed | 20-50 mm/s |
| Scanable range (metal-free) | 160 cm |
| Examination room Temperature range | 18–28° C |
| Heat dissipation air-cooled | Max. 12 kW |
| Heat dissipation water-cooled | Max. 12 kW |
| Relative air humidity (no condensation) | 20-75% |
| Surface area for installation | 18 m2 (24 m2 incl. equipment room) |
| Generator power | 100 kW |
| Nominal line frequency | 50; 60 Hz |
| Max. power connection | <70 kVA |
| Power consumption System on standby | <4.0 kVA |
| Power consumption scanning | 125 kVA (at 80 kW) 140 kVA (at 100 kW) |
| Nominal voltage 3/N~ | 380-480 V in 20 V steps |