Zustand: Wiederaufbereitet durch MEDSER, hervorragender technischer und optischer Zustand
Standort: MEDSER Showroom, Heusenstamm/Frankfurt
Röhre: Dura 422 MV
Verfügbarkeit: Sofort lieferbar, Vorverkauf vorbehalten
| Aperture | 70 cm |
| Scan Field | 50 cm |
| Tilt | +/- 30° |
| Rotational times | 0.6, 1.0, 1.5 s |
| Number of detector rows | 6 |
| Elements | 11.776 |
| Channels per slice | 1.472 |
| Number of projections | Up to 1.250 (1/360°) |
| Max. slices / rotation | 6 |
| Reconstruction time | up to 8 images/s |
| Reconstruction slice widths | 0.63, 0.75, 1.0, 1.25, 2.0, 2.5, 3.0, 4.0, 5.0, 6.0, 8.0, 10.0 mm |
| Slice increment | 0.1–10 mm |
| Pitch factor (Volume Pitch) | 0.4 – 1.8 |
| Scan length | max. 150 cm |
| Spiral scan time | Max. 100 s |
| Tube | DURA 422MV |
| Tube current range | 20–345 mA |
| Tube voltages | 80, 110, 130 kV |
| Anode heat storage capacity | 5 MHU |
| Focal Spot size according to IEC 60336 | 0.8 x 0.5 mm/7° 0.8 x 0.7 mm/7° |
| Max. table load | 200 kg |
| Table speed | 1–100 mm/s |
| Vertical table travel range | 45–83 cm (at table top) |
| Vertical travel speed | <22,4 mm/s |
| Scanable range (metal-free) | 153 cm |
| Examination room Temperature range | 18–30° C |
| Heat dissipation gantry | Max. 6.8 kW |
| Surface area for installation | 18 m2 |
| Power Supply | 380-480 V |
| Nominal line frequency | 50; 60 Hz |
| Max. power connection | <70 kVA |
| Power consumption System on standby | <3,7 kW |
| Power consumption scanning | <7,0 kW |
2D post processing
Specially designed X-ray exposure filter installed at the tube collimator. Up to 25% dose reduction with increased image quality.
Evaluation of spiral images and display of vessels, vascular anomalies, aneurysms, plaques and stenosis
Interface for transmitting medical images and information in the DICOM industrial standard. Permits communication between devices from different manufacturers.