Baujahr: 2013
Zustand: wiederaufbereitet durch MEDSER
Standort: MEDSER Showroom, Heusenstamm/Frankfurt
Röhre: Straton MX-P46
Aperture | 78 cm |
Scan Field | 50 cm |
Tilt | +/- 30° |
Rotational times | 0.33, 0.5, 1.0 s |
Number of detector rows | 32 |
Elements | 23.552 |
Number of projections | Up to 4.608 (1/360°) |
Max. slices / rotation | 64 |
Field of view | 50, 65 HD, 80 Extended cm |
Reconstruction time | up to 40 images/s |
Reconstruction slice widths | As low as 0.33 mm |
Slice increment | 0.1–10 mm |
Pitch factor (Volume Pitch) | Up to 1.5 |
Scan length | max. 160 cm |
Spiral scan time | Max. 100 s |
Tube | STRATON |
Tube current range | 20–800 mA |
Tube voltages | 70, 80, 100, 120, 140 kV |
Anode heat storage capacity | 50 MHU (cooling rate of 7.3 MHU/min) |
Focal Spot size according toIEC 60336 | 0.7 x 0.7 mm/7° 0.9 x 1.1 mm/7° |
Max. table load | 220 kg |
Table speed | 2–200 mm/s |
Vertical table travel range | 48–92 cm (at table top) |
Vertical travel speed | 20-50 mm/s |
Scanable range (metal-free) | 160 cm |
Examination room Temperature range | 18–28° C |
Heat dissipation air-cooled | Max. 12 kW |
Heat dissipation water-cooled | Max. 12 kW |
Relative air humidity (no condensation) | 20-75% |
Surface area for installation | 18 m2 (24 m2 incl. equipment room) |
Generator power | 100 kW |
Nominal line frequency | 50; 60 Hz |
Max. power connection | <70 kVA |
Power consumption System on standby | <4.0 kVA |
Power consumption scanning | 125 kVA (at 80 kW) 140 kVA (at 100 kW) |
Nominal voltage 3/N~ | 380-480 V in 20 V steps |