Baujahr: 2003, refurbished by SIEMENS® 2010
Zustand: wiederaufbereitet durch MEDSER
Standort: MEDSER Showroom, Heusenstamm/Frankfurt
Röhre: Dura 532, Akron B
Aperture | 70 cm |
Scan Field | 50 cm, 25 cm |
Tilt | +/- 30° |
Rotational times | 0.5, 0.75, 1.0, 1.5 s |
Number of detector rows | 8 |
Elements | 5376 |
Channels per slice | 1.344 |
Number of projections | up to 2.320 (1/360°) |
Max. slices / rotation | 4 |
Reconstruction time | 1.5 slices/s |
Reconstruction slice widths | 0.5–10.0 mm |
Slice increment | 0.1–10 mm |
Pitch factor (Volume Pitch) | 1–8 |
Scan length | max. 157 cm |
Spiral scan time | Max. 100 s |
Tube | DURA Akron B |
Tube current range | 28–580 mA |
Tube voltages | 80, 120, 140 kV |
Anode heat storage capacity | 5.3 MHU |
Focal Spot size according to IEC 60336 | 0.5 x 0.7 mm/7° 0.8 x 1.2 mm/7° |
Max. table load | 200 kg |
Table speed | 1–150 mm/s |
Vertical table travel range | 48–102 cm (at table top) |
Vertical travel speed | 2.5–50 mm/s |
Scanable range (metal-free) | 157 cm |
Examination room Temperature range | 15-28°„ C |
Heat dissipation gantry | max. 13,5 kW |
Surface area for installation | 30 m2 |
Power supply | 380-480 V |
Nominal line frequency | 50; 60 Hz |
Max. power connection | 60–83 kVA (fuse 100 A) |
Power consumption Computer on | 4 kVA, 1.1 kW |
Power consumption System on standby | 7 kVA, 5 kW |
Power consumption Scanning | 48 kVA |
Interface for transmitting medical images and information in the DICOM industrial standard. Permits communication between devices from different manufacturers.